71 • Appendix E: Infrared Methane Sensors

Eagle Instruction Manual

case. It is connected to the EC4 socket on the analog PCB.

4. Adjust the sensor’s SPAN control one turn and observe the

display reading. Continue to adjust the SPAN control until the
display reading matches the concentration of the calibration
cylinder.

The SPAN control is on the top edge of the amplifier and is the
control closest to the front of the instrument.

5. Place the top of the case in its original position, then secure it with

the large screws you loosened in step 5.

6. Continue with the normal calibration procedure.

Replacing Infrared Methane Sensors

Return the Eagle to RKI Instruments, Inc. for replacement of the
infrared methane sensor when:

• The infrared methane channel cannot be calibrated correctly.

• The infrared methane reading cannot be set to zero by the Demand

Zero command.

Parts List

Table 15 lists part numbers for accessories needed to calibrate the
Eagle’s infrared 100% volume methane sensor.

Table 15: Parts List: Infrared Methane Sensors

Part Number

Description

81-0015RK-01

Calibration cylinder, 34-liter (50% volume CH

4

)

81-1001RK

Dispensing valve (for 34-liter calibration cylinders)

81-1126RK

Calibration sample bag, tedlar (9 in. x 9 in.)

81-1127RK

Calibration sample bag, tedlar (12 in. x 12 in.)