Eagle Instruction Manual

Setup Mode • 30

Updating the Display Increment Setting

1. Press the SHIFT/▼ button to place the prompt in the third line,

then press the POWER/ENTER button. The display increment
setting flashes. The allowable settings are 0.2 VOL% (default) and
0.5 VOL%.

2. Press the AIR/▲ or SHIFT/▼ button to display the desired

display increment setting, then press the POWER/ENTER button
to enter the setting. The prompt flashes.

Returning to the Main Menu

1. Press the SHIFT/▼ button. The

ESCAPE

message displays. (Press

the AIR/▲ button to return to the previous screen.)

2. Press the POWER/ENTER button to save the new settings. The

OTHER GAS SET

message displays, then the Gas Combinations

menu displays.

3. To exit the Gas Combinations menu, press the SHIFT/▼ button

until the prompt is next to Channel 4, then press the SHIFT/▼
button again. The

ESCAPE

message displays.

4. Press the POWER/ENTER button. The message

SAVING DATA

displays, then the main menu displays.

Updating Toxic Channel Settings

This section describes how to update the target gas label, set a custom
gas label, and update the fullscale and display increment settings for a
toxic gas channel.

Updating the Target Gas Label

1. From the main menu, select the GAS COMBINATIONS menu

option.

2. Press the POWER/ENTER button to display the Gas

Combinations menu.

3. Use the AIR/▲ or SHIFT/▼ button to place the prompt next to the

toxic gas channel (in this example H2S or CO).

4. Press the POWER/ENTER button. The toxic target gas label

flashes. This indicates that this setting can now be updated.

C H

4

O

Y

C O

H

2

S

X

>