RKI Instruments Eagle 1 Manual User Manual

Page 36

Advertising
background image

29 • Setup Mode

Eagle Instruction Manual

Updating Oxygen Channel Settings

This section describes how to update the target gas label, fullscale
setting, and display increment setting for the oxygen channel.

Updating the Target Gas Label

1. From the main menu, select the GAS COMBINATIONS menu

option.

2. Use the AIR/▲ or SHIFT/▼ button to place the prompt next to the

oxygen channel (in this example OXY).

3. Press the POWER/ENTER button. The oxygen target gas label

flashes. This indicates that this setting can now be updated.

4. Press the AIR/▲ or SHIFT/▼ buttons to display the available

oxygen target gas labels (OXY, ***, and NOT USED).

CAUTION:

The *** setting is not intended for customer setup. Contact RKI
Instruments, Inc. before using this setting for the oxygen
channel.

5. Press the POWER/ENTER button to enter the new target gas label.

6. To exit the Gas Combinations menu, press the SHIFT/▼ button

until the prompt is next to Channel 4, then press the SHIFT/▼
button again. The

ESCAPE

message displays.

7. Press the POWER/ENTER button. The

SAVING DATA

message

displays, then the main menu displays.

Updating the Fullscale Setting

1. Press the SHIFT/▼ button to place the prompt in the second line,

then press the POWER/ENTER button to update the fullscale
setting. The fullscale setting flashes.

The maximum fullscale setting for the oxygen channel is 40.0
VOL%; the minimum setting is 25.0 VOL%. The default setting is
40.0 VOL%.

2. Press the AIR/▲ and SHIFT/▼ buttons to display the desired

fullscale setting, then press the POWER/ENTER button to enter
the setting. The prompt flashes.

C H

4

O

Y

C O

H

2

S

X

>

Advertising