Replacing the Infrared HC Sensor

Return the Eagle to RKI Instruments, Inc. for replacement of the
infrared HC sensor when:

• The infrared HC channel cannot be calibrated correctly.

• The infrared HC reading cannot be set to zero by the Demand Zero

command.

Parts List

Table 20 lists part numbers for replacement parts and accessories of
the Eagle’s infrared HC sensor.

Table 20: Parts List: Infrared HC Sensors

Part Number

Description

06-1248RK-03

Tubing, 3/16 x 5/16, polyurethane, 3 foot length, for calibration kit

81-0018RK

Calibration cylinder, 50% LEL isobutane in air, 17 liter steel

81-0018RK-01

Calibration cylinder, 50% LEL isobutane in air, 34 liter steel

81-0018RK-03

Calibration cylinder, 50% LEL isobutane in air, 103 liter steel

81-1001RK

Dispensing valve (for 34-liter calibration cylinders)

81-1050RK

Regulator with gauge and knob, 0.5 LPM, for 17 liter and 34 liter
steel calibration cylinders

81-1051RK

Regulator with gauge and knob, 0.5 LPM, for 34AL/58/103 liter
calibration cylinders

81-0158RK-02

Four-gas calibration cylinder, 50% LEL isobutane/12% O

2

/50 ppm

CO/25 ppm H

2

S, 58 liter aluminum

81-0158RK-04

Four-gas calibration cylinder, 50% LEL isobutane/12% O

2

/50 ppm

CO/25 ppm H

2

S, 34 liter aluminum

81-1126RK

Calibration sample bag, tedlar (9 in. x 9 in.)

81-1127RK

Calibration sample bag, tedlar (12 in. x 12 in.)