57 • Appendix C: Non-Standard Toxic Gas Sensors

Eagle Instruction Manual

WARNING: The Eagle’s alarms are initiated by the DISPLAY reading

not the FACTORED reading. If you are monitoring for
toluene as in the above example and the low alarm is set
for 10% LEL, the Eagle will initiate a low alarm at 7%
LEL toluene (display reading of 10% LEL).

Appendix C: Non-Standard Toxic Gas Sensors

Appendix C describes the Eagle’s non-standard, electrochemical toxic
gas sensors. This appendix also offers information for calibrating and
replacing non-standard toxic gas sensors. Non-standard electro-
chemical toxic gas sensors are toxic gas sensors other than CO or H

2

S.

Specifications

Table 8 lists specifications for the non-standard toxic gas sensors. The
alarm settings are user-adjustable (see “Updating the Alarm Point
Settings” on page 33).

Table 8: Non-Standard Toxic Gas Sensors Specifications

Target Gas

Range

Alarm 1

Alarm 2

TWA

Alarm

STEL

Alarm

Ammonia

0 to 75.0 ppm

12.0 ppm

25.0 ppm

25.0 ppm

35.0 ppm

Arsine

0 to 1.00 ppm

0.05 ppm

0.10 ppm

0.05 ppm

OFF

Arsine

0 to 200 ppb

1

50 ppb

100 ppb

50 ppb

OFF

Bromine*

0 to 1.00 ppm

0.05 ppm

0.10 ppm

0.10 ppm

OFF

Chlorine

0 to 3.00 ppm

0.50 ppm

1.50 ppm

0.50 ppm

1.00 ppm

Chlorine
Dioxide*

0 to 1.00 ppm

0.10 ppm

0.30 ppm

0.10 ppm

0.30 ppm

Diborane

0 - 300 ppb

1

50 ppb

100 ppb

100 ppb

OFF

Fluorine**

0 to 5.00 ppm

0.50 ppm

1.00 ppm

1.00 ppm

2.00 ppm

Hydrogen
Bromide*

0 to 9.00 ppm

1.00 ppm

2.00 ppm

OFF

3.00 ppm

Hydrogen
Chloride*

0 to 15.0 ppm

2.0 ppm

5.0 ppm

OFF

OFF

Hydrogen
Cyanide

0 to 30 ppm

3.0 ppm

5.0 ppm

OFF

4.7 ppm

Hydrogen
Fluoride**

0 to 9.00 ppm

1.00 ppm

2.00 ppm

1.00 ppm

2.00 ppm