Page 24

Short-Term Exposure Limit (STEL) Screen

The STEL function shows the average reading for H

2

S and CO during

the last 15 minutes. These values are not affected by the Lunch Break
function and are reset every time the unit is turned off.

Time-Weighted Average (TWA) Screen

The TWA function shows the average reading for toxic gases during the
last 8 hours
. If 8 hours has not elapsed since the last time the STEL/
TWA was cleared, the average is still calculated over 8 hours, with the
missing time assigned a 0 value for readings.

When the Lunch Break RESUME option is selected during start-up, the
GX-94 will include the TWA from the last session.

STEL

H2S 0.

0

PPM

CO 0PPM

TWA

H2S 0.

0

PPM

CO 0PPM