Auto cal. <hc> 50 lel – RKI Instruments GX-94 User Manual

Page 37

Advertising
background image

GX-94 Operator’s Manual

Page 36

one or more concentrations do not match, continue with step 4.

4. To adjust the values on this screen, hold down the (SHIFT)/

but-

ton and press the DISPLAY/(ADJ.) button. The AUTO CAL. value
screen for the combustible gas channel displays.

5. Use the AIR/

(increase) and (SHIFT)/

(decrease) buttons to

set the correct combustible gas value.

6. Press the ON/OFF/INPUT button to enter the new setting. The

AUTO CAL. value screen for the next channel displays.

7. Repeat steps 5 and 6 to set the correct values for the remaining

channels and return to the Calibration Values screen.

Note

The RKI Four Gas Cylinder contains approximately 12% O

2

by volume. Be sure to set the “O2” auto calibration value to
agree with the concentration listed on the cylinder’s label, not
zero.

8. With the Calibration Values screen displayed, press the ON/OFF/

INPUT button. The gas readings will flash.

9. Screw the regulator onto the calibration cylinder. The gas begins to

flow.

10. Allow the gas to flow for one minute (the readings should be sta-

ble) then press the ON/OFF/INPUT button to set the calibration for
each channel to the programmed values.
If a sensor(s) cannot be calibrated to the proper value, FAIL PUSH
AIR
displays and the GX-94 lists the sensor(s) that failed to

AUTO CAL.

<HC>

50 LEL