Auto cal 2.set span <hc – RKI Instruments GX-94 User Manual

Page 38

Advertising
background image

Page 37

GX-94 Operator’s Manual

calibrate. The other sensors calibrate normally. The buzzer and
alarm LED activate. Press the AIR/

button to reset the alarm and

return to the main menu. Replace the failed sensor(s), then repeat
calibration.

11. AUTO CAL END displays, then the Calibration menu displays.
12. Unscrew the regulator from the cylinder.
13. Remove the adapter plate from the GX-94.
14. Press the SHIFT/

button to place the prompt next the 7.RETURN

menu option, then press the ON/OFF/INPUT button to return to
normal operation.

Calibrating with the Single Calibration Method

This section describes calibration using the single calibration method.
To calibrate using the auto calibration method, see Calibrating with
the Auto Calibration Method
earlier in this section.

CAUTION

The single calibration method does not have a “FAIL”
notification. Replace sensors that cannot be set to agree
with the calibration source, then recalibrate.

1. Use the AIR/

and (SHIFT)/

button to place the cursor next to

the channel you want to calibrate. In this example the combustible
gas channel is calibrated. Place the cursor next to the SET SPAN
<HC> menu option. The number 2 will flash.

:

1.AUTO CAL

2.SET SPAN
<HC>