Auto cal 2.set span <hc, Setting the date and time – RKI Instruments GX-94 User Manual

Page 40

Advertising
background image

Page 39

GX-94 Operator’s Manual

reads “OXYGEN ZERO”. Verify the concentration of oxy-
gen listed on the cylinder’s label and set the zero reading
to the same concentration. A 100% nitrogen source (0%
oxygen) is recommended when using single calibration.

7. After the last channel is calibrated, press the (SHIFT)/

button

until the number 7 in front of RETURN, the last menu item,
flashes.

8. Press the ON/OFF/INPUT button to return to normal operation.

Setting the Date and Time

The number 6 option in the calibration menu is SET DATE/TIME. Use
this item to set the current date and time for the GX-94.

From the normal operation screen, press and hold the (SHIFT)/

but-

ton, then press the DISPLAY/(ADJ.) button to enter Calibration Mode.
The calibration menu will show the following screen with a flashing
number indicating the cursor position:

1. Use the (SHIFT)/

button to place the cursor in from of menu item

6. SET DATE/TIME. The number 6 will flash.

2. Press the ON/OFF/INPUT button. The display will show the cur-

rent settings, with the month flashing.

2. Press AIR/

or (SHIFT)/

to scroll to the desired month.

3. Press ON/OFF/INPUT to enter the selection and go to the next

setting (day).

4. Repeat steps 2 and 3 for each value: day, year, hours, minutes.

You may skip past a correct setting by pressing ON/OFF/INPUT.

1.AUTO CAL

2.SET SPAN
<HC>

Advertising