Updating the Gas Combination (2. COMBINE GAS)

Sets the display and circuitry for types of gases to be detected; there
are 11 possible combinations. Press AIR/

▲ or

(SHIFT)/

to scroll

through the combinations. Press ON/OFF/INPUT to accept the selected
combination.

Note

Change the GAS COMBINATION values only when the
appropriate sensors are installed in the corresponding sensor
sockets. Calibrate immediately after changing sensor combi-
nations.
Gases selected in this operation will be displayed throughout
all setup and normal display screens.

Viewing & Setting Alarms (3. SET ALARM)

Sets the alarm levels for gases detected by the GX-94. Press AIR/

to

change the alarm level for a single gas, or DISPLAY/(ADJ) to skip that
channel.

Press AIR/

to raise an alarm level, or (SHIFT)/

to lower the alarm

level. Hold down the button to scroll quickly through the alarm setting.
The lowest available setting for each alarm point is “off.”

Press ON/OFF/INPUT to accept the displayed alarm value and move to
the next alarm or gas.

SET ALARM <HC>

Allows the user to change LOW and HIGH HC alarms for
combustible gas.

SET ALARM <O2>

Allows the user to change the LOW and HIGH %O

2

alarms.

Remember that the LOW O

2

alarm is a decreasing alarm, for falling