Entering setup mode, Resetting the nvram (1. set def nvram), Def. set all – RKI Instruments GX-94 User Manual

Page 29: Def. set alarm, Def. set o2 zero, Def. set temp. zero

Advertising
background image

GX-94 Operator’s Manual

Page 28

AIR/

and (SHIFT)/

buttons.

Entering Setup Mode

1. Take the GX-94 to a fresh air location—the instrument does not

detect gas while in setup mode.

2. If it is not already off, turn the GX-94 off.
3. Hold down the AIR/

and (SHIFT)/

buttons simultaneously, then

press ON/OFF/INPUT. The main menu will appear (release the
buttons at this time). The “1” will be flashing indicating the cursor
position and that this item is currently selected.

4. Scroll to the desired main menu item using the AIR/

and

(SHIFT)/

buttons and press ON/OFF/INPUT to enter that item.

Resetting the NVRAM (1. SET DEF NVRAM)

DEF. SET ALL

Allows the user to reset all values for non-volatile RAM; all options
are returned to default values.

DEF. SET ALARM

Resets only the alarm levels to default values.

DEF. SET O2 ZERO

Resets only the O

2

zero setting to the default value.

DEF. SET TEMP. ZERO

Resets only the temperature zero setting to the default value.

1.SET DEF
NVRAM
2.COMBINE
GAS