10

3−4.Flow monitor check

When the power gets ON, the pump shall start operation check that suction works properly by
filter tube with flow monitor.

At normal sucking(Can see float)


Not sucking(Can not see float)


3−5.Zero point adjustment

Zero point calibration for this instrument shall be done HC, O2(Oxygen), H2S(Hydrogen
sulfide).





Float

Caution

When make zero adjustment, proceed to do it after having fresh air sucked into instrument
for over 1 min. When make zero adjustment through the suction of gas present atmosphere,
the correct measurement can not be done and it may lead to the accident.