24

6−2.Gas calibration (Span adjustment)






ZERO key

PUMP key

・ Hold and pressing ZERO and PUMP key

similtaneously in the operation mode.


・ First, CH4 display shall be shown.
・ When press EMTER key, “HC” gas

calibration shall start.

・ When press 『+』key, it shall change to

gas calibration of 「OXY」→「H2S」.

・ When pressing ESC key, it shall return to

“Operation display”.

HOLD KEY

SPAN GAS HC

ENTER key

ESC key

After displaying for 5 sec,

proceed to the next display.

・This shall display gas name, gas density and unit.

( When press ESC key, it shall return to HC

display.)

・Gas density shall blink.

・ Have cal gas sucked into instrument and unit the

reading gets stable.

・When the gas density differs, adjust to cal gas

value by 『+』or『−』 key.

・After adjusted to it, fax the span by pressing

ENTER key.

・ After span adjusted, proceed to the oxygen

display.

・ For gas calibration of O2 and H2S make the

same operation.

・ After completion calibration, go back to

the operation display by pressing “ESC

key”.

SPAN GAS OXY

ESC key

《SPAN cal OK》

《Span cal fails》

A D J > −/+

S P A N S E T O K

F A I L S P A N

ENTER key

SPAN GAS H2S

*Note

《RX-517》

・For measurement of H2S high, the mode display shall be changed from “H2S high” to “cal

display”.

・When mode switch is ON during the cal mode the cal mode shall be cancelled and changed

to operation mode.