Electro-chemical cell method – RKI Instruments RX-517 User Manual

Page 35

Advertising
background image

PT0E-0730

34


11−3.ELECTRO-CHEMICAL CELL method

This gas detector applies an electrochemical sensor. The sample gas is electrolyzed by the
electrolyzed cell added with specific electric potential (bias voltage) and detected from the
electrolyzed current generated at that time.
The electrochemical sensor is designed to keep the interface between electrode and electrolyte
at a constant potential (Bias voltage) and is the method to electrolyze gas directly. Then as
the gas has the bias voltage generating its own electrolization (Rrdox reaction) the bias voltage
of sensor is determined by the redox potential.

32

電解液

R(抵抗)

参照電極

対極

作用電極

気体透過膜

Gas in

Gas out

構造図

ポテンショスタット回路

PS circuit

RE

CE

GPF

WE

Figure

Solution

Advertising