5−1−1.Alarm setting


1

st

alarm display

PEAK key

M ‐ P A S S W O R D

・First, 1

st

alarm point shall be shown on the

display.

・Change 『+』or『−』key and confirm it by

pressing ENTER key.

・When press ESC key, the setting shall be void

and first alarm point shall be shown.

HC:1

st

alarm point

HC:2

nd

alarm point

OXY:1

st

alarm point

OXY:2

nd

alarm point

H

2

S:1

st

alarm point

H

2

S:2

nd

alarm point

High H

2

S:1

st

alarm point(RX-517)

High H

2

S:2

nd

alarm point(RX-517)

Adjust it

in this

procedure.

ENTER key

ESC key

ENTER key

ENTER key

SAVE?

END

W A R N I N G

ESC key

ESC key

ESC key

ESC key

ESC key

・ After finishing 2

nd

alarm of H

2

S is

adjusted, it displayed “SAVE ?”.

W A R N I N G

W A R N I N G

A L A R M

・Input the password【3002】.

・When press『+』or『−』, the number will be

changed.

・Fix the number by ENTER key and move to next

digit.

・After fixing the last digit, move to alarm

setting (When make the password error, it will

show “ERROR”. Press ESC key.)

・Then 2

nd

alarm point of CH4 shall be shown on

the display.

・When change by『+』or『−』key, confirm it

by ENTER key.

・When press the setting by ESC key, go back to

1

st

alarm point.

・When press “ENTER” key, save the set value

and shall show “END” display and go back

to 1

st

alarm point display.

・When press “ESC” key, the set value is

canceled and go back to 1

st

alarm point

display.