《Interference measures in inert gas》

In the component of inert gas, a lot of interference gas may be present. When it should be
present so much to affect the measurement, it shall be required to mount the interference
removal filter.

4.Alarm

4−1.Alarm display

①At 1

st

alarm, the figure will blink and “Warring” will blink.

②At 2

nd

alarm, the figure will blink and “Alarm” will blink.

③At over alarm, the figure will blink and “Over” will blink.

※Non latched mode after alarm


4−2.Alarm buzzer & lamp action

《2

nd

alarm》

《Over alarm》

W A R N I N G

A L A R M

O V E R

Alarm≧1

st

alarm set(HC、H

2

S)

Alarm≦1

st

alarm set(O

2

【WARNING display, Intermittent LED, Blink】

Alarm≧2

nd

alarm set

【ALARM display, Intermittent LED, ON】

Alarm≧F.S.

【OVER display, Intermittent LED, ON】

《1

st

alarm》

WARNINNG

(1

st

alarm)

Lamp

ALARM,OVER

(2

nd

alarm, Over alarm)

Lamp

【Intermittent tone】

【LED blinking】

【Continuous tone】

【LED light-on】