《Cal gas make-up by H2S gas checker CK-82》

Take out 1 pce. ampoule from accessory ampoule pack.

Check that the rubber caps are fixed in the cover of container.

Open the cover of container and deflate the air by squeezing the pressure adj bag.

Put up the ampoule crusher know and put the ampoule into ampoule crusher inside of the
cover shown below. When the ampoule is put in reversely, the ampoule may be crushed.










After the ampoule is put in, close the cover.

Check that the rubber caps are put on the top cover.

After checking that the cover is closed, put in the ampoule crusher knob and crush the
inside ampoule.

Grip the container with hands and shake it to up & down left & right slightly so that
H2S can not remain in the ampoule crushed.

About 21 ppm H2S is supposed to be produced in the container.























Caution

・21ppm H2S gas density by CK-82 is the theoretical value in the calculation. Actually,

the error of 1-2ppm could be expected. When desired to check H2S in the container,
check it by detector tube. When use gas checker CK-82, see the separate manual for
model CK-82 together.

Ampoule

Ampoule crusher knob

Crush the ampoule by pressing the ampoule crusher knob