RKI Instruments RX-517 User Manual

Page 12

Advertising
background image

PT0E-0730

11

<FLOW CHART OF ZERO POINT ADJ>

ZERO MODE DISPLAY

HOLD KEY

ADJ ZERO OK

《Zero set OK》

ZERO SET OK

FAIL ZERO

・Press ZERO key for 3 sec.

Press for 5 sec and move to

the display.

《Zero adj fails》

FRESH AIR?

・When press ENTER key, HC, O2, H2S zero adj

shall start.

・Calibration

ZERO key

ENTER key

ESC key

Move to zero adj

for CO2 adj ESC key

All are OK

HC fails

*Note

・To cancel previous operation, press ESC key.
But can not cancel after press ENTER key.

《RX-517 case》
・For zero adjustment of H2S high, proceed to make zero adjustment by changing to the

display of H2S high