PT0E-0730

13

3−7.Peak SW value mode

When press PEAK SW during operation, the peak valve from power on to the current point (HC,
H2S = MAX, O2=MIN) shall be shown.

When hold and press ESC key for 3 sec during the peak value display, the peak value shall be
cleared.





3−8.Power pump ON/OFF

When move the sampling point and stop to operate temporarily, the pump only stops operation.
When stop the pump operation, the power consumption shall be reduced by about 40%. Then
the other than pump is on active and once the pump re-starts up, this shall start operation
immediately.




Peak mode

Operation mode

ESC SW

P E A K

Operation mode

Peak mode

PEAK SW

P E A K

【RX-517】

H2S change over

P E A K

H2S high display

Peak display

*Note

《RX-517 case》
・In the H2S high display, the peak value of H2S is displayed.
・When changed from peak display, it changes to the operation display.

【RX-516/517】

PEAK key

*Note

・When do not operate for the time being, make off the power. Other than pump is active

and if it is left alone, the battery is consumed up.

*注記

《RX-517 case》

All peak value is cleared on both the displays of HC, O2, H2S low and the another one
of H2S high.