RKI Instruments RX-517 User Manual

Page 24

Advertising
background image

PT0E-0730

23

6.Maintenance check

For this correct operation, follow to the following procedure.




6−1.Battery replacement








When replace the batteries, replace all four pcs with new ones.

① Check that the power if off. But when the power is on, make off the power first by use

of POWER/ENTER key.

② Remove the carrying case from the unit.
③ Open the battery cover on the bottom of unit by turning the screw with (-) screwdriver

or coin etc counter-clockwise.

④ Remove 4 pcs batteries and put on new batteries by taking care of battery polarity.
⑤ Close the battery cover.










Warning

・ For battery replacement, make it non-hazardous area by all means after turning off the

power.

・ Be sure to change the battery with the genuine ones.

Caution

・Bure sure to close the battery cover.

Battery cover

*Note

・When remove the battery, remove it from (+). When put on, do it from (-) polarity.

As this is safety instrument. Be sure to make the regular maintenance. If any trouble
should take place, contact RIKEN KEIKI or nearest agent.