14 • Setting the Bump Test and Calibration Parameters

Air Sample Time (Air)

The air sample time can be set separately for bump testing and
calibration. It is the length of time that a SM-2009U will draw air through
the AIR fitting on the back of the station. Air is drawn during a bump test
before an air adjust operation and to purge calibration gas from the
system after calibration gas has been drawn through the CAL. GAS fitting
on the back of the station.

Calibration Gas Sample Time (GAS)

The calibration gas sample time can be set separately for bump testing
and calibration. It is the length of time that a SM-2009U will draw

Table 3: Bump Test & Calibration Parameter

Parameter

Display

Tag

Available

Choices

Bump

Factory

Setting

Cal

Factory

Setting

Air Sample
Time

Air

• 30 seconds
• 45 seconds
• 60 seconds
• 90 seconds

30 seconds

45 seconds

Calibration Gas
Sample Time

GAS

• 30 seconds
• 45 seconds
• 60 seconds
• 90 seconds
• 120 seconds

30 seconds

90 seconds

Bump Test
Check Toler-
ance

cHE

• ± 10%
• ± 20%
• ± 30%
• ± 40%
• ± 50%

50%

n/a

Automatic Cali-
bration

cAL

• On
• Off

Off

n/a