Setting the Bump Test and Calibration Parameters • 17

NOTE:

You can still use the SM-2009 with an instrument for bump
testing and calibration if this failure indication occurs, but you
will not be able to perform any operations that require the
display until a calibration is performed, the unit is turned off,
and then the unit is re-connected to the SM-2009U. If you
perform a bump test or calibration in this case, the only
indication of pass or fail will be the BUMP or CAL. LED.

When Cal Limit Display is set to On and Cal. Limit Check is
set to Confirm to use, if the GX-2009 is due for calibration the
unit will show the following screen and beep and flash the LED
arrays for a few seconds indicating that calibration is due.

When the beeping and flashing stops, press and release the
POWER MODE button on the GX-2009 to continue. The unit will
now connect to the SM-2009U, the TRANSMIT screen shown in
step 6 above will be displayed, and the CHARGE LED will begin
to blink green.

Setting the Bump Test Parameters

Do the following to set the bump test parameters after establishing a
connection between a GX-2009 and the SM-2009U.

1. Turn on the SM-2009U with a GX-2009 and establish a connection

between them as described above in "Turning On the SM-2009U
with a GX-2009". The GX-2009 will display the TRANSMIT screen.

CAL

C--LIMIT

50 12.0

TRANSMIT

CO ppm H2S ppm

CH4 %LEL OXY vol%

50 25.0