16 • Setting the Bump Test and Calibration Parameters

5. Install a GX-2009 in the instrument cradle. The CHARGE LED will

begin to blink yellow. See “Charging an Instrument in a Calibration
Station” on page 35 for a complete description of charging the GX-
2009.

6. Press and hold the POWER MODE button on the GX-2009 until

you hear a beep, then release it. The GX-2009 will begin it’s power
up sequence (see exceptions below in step 7). If a successful
connection between the GX-2009 and the SM-2009U occurs, the
TRANSMIT screen will appear on the GX-2009 display.

NOTE:

The screen shown above applies to a 4-channel GX-2009. If
your GX-2009 has less than 4 channels, the inactive
channels will not appear in the above screen and in any
screens where channels are displayed. All screens in this
manual assume a 4-channel GX-2009.

7. There are two exceptions to the sequence described in step 6

above. See the GX-2009 User Setup Program Operator’s Manual
for a description of the Cal Limit Display and Cal Limit Check
instrument parameters.

When Cal Limit Display is set to On and Cal. Limit Check is
set to Can’t Use, if the GX-2009 is due for calibration the
instrument will show a calibration expired failure screen and the
buzzer and LED arrays will pulse for a few seconds before
connecting to the SM-2009U.

The failure screen will remain on the display while the unit is
connected to the SM-2009U and it is not possible to view or set
the bump test or calibration parameters.

50 12.0

TRANSMIT

CO ppm H2S ppm

CH4 %LEL OXY vol%

50 25.0

FAIL

C-LIMIT