Calibration and Bump Test Records • 37

it, the CHARGE LED will continue to blink yellow while charging is
taking place. The SM-2009U will take approximately 3 hours to
charge a fully discharged GX-2009.

6. When the charge is complete, the CHARGE status LED will turn

solid green.

Calibration and Bump Test Records

The SM-2009U saves a record of each bump test and calibration
performed. It is capable of saving up to 200 such records. When an SM-
2009U’s memory becomes full, the oldest record is overwritten when a
new record is saved. The records saved in the SM-2009U’s memory can
be saved to a USB flash drive using the USB port on the front.

Available Memory in the SM-2009U

The COPY LED indicates how much of the SM-2009’s memory has been
used. The table below describes the various indications.

Copying Calibration and Bump Test Records to a USB
Flash Drive

The SM-2009U will only perform a copy operation if there is at least one
record saved in it’s memory. If there are no records saved in the station’s
memory, the COPY LED will be off and the COPY button will not function.
Do the following to save calibration and bump test records in the
SM-2009U’s memory to a USB flash drive.

Table 5: COPY LED Indications

COPY LED Indication

Memory Used

Off

None. No records are saved

Solid Green

Less than 80% of the SM-2009U’s
memory has been used.

Solid Yellow

More than 80% of the SM-2009U’s
memory has been used.

Solid Red

The calibration station’s memory is
full. A newly saved record will over-
write the oldest one.