RKI Instruments Beacon 200 User Manual

Page 13

Advertising
background image

Internal Description • 13

AC In Terminal Strip. The AC in terminal strip is a 3-point terminal strip
located above the controller terminal strip (see Figure 1). It facilitates
wiring from the AC power source. Table 4 lists the function of each
terminal.

Relays

The Beacon 200 includes four channel relays (two per channel) and three
common relays. Both sets of relays are single-pole, double-throw (SPDT)
and are rated for 10 amps at 250 VAC (resistive).

NOTE: You can select normally energized (NE) or normally de-energized

(NDE) settings for each channel in the Channel Control and Setup
program. This section describes the default setting: normally de-
energized.

The alarm 1 and alarm 2 common relays are factory-set as NDE
and the fail common relay is factory-set as NE. The alarm 1, alarm
2, and fail common relays’ NE/NDE settings are not user-
adjustable.

Channel relays. The four channel relays are above the alarm terminal
strip (see Figure 1). These relays are dedicated to specific channels
and alarm levels.

For example, the channel 1, alarm 1 relay energizes if channel 1
recognizes an alarm 1 condition. Figure 3 below illustrates the
allocation of the channel relays

Table 4: Terminal Assignments for the AC In Terminal Strip

Terminal

Connects to:

LINE

Hot wire from AC power source.

NEUT

Neutral wire from AC power source.

GND

Earth ground