Setting Up a New Channel or Changing an Existing Channel • 43

When the desired full scale setting appears, press the ENTER button
to accept the setting. The save screen appears.

One of the choices is User Will Specify. If the desired full scale setting
is not in the list, this setting will allow you to enter a full scale setting.
With this choice displayed, press the ENTER button and a screen will
appear which prompts you to choose how many decimal places you
want in the full scale setting. You can select up to 3 decimal places,
then press the ENTER button to proceed to the select full scale screen.
Use the UP/YES and DOWN/NO buttons to select a full scale setting,
then press ENTER to accept it. The save screen appears.

11. To save the settings, press the UP/YES button and the settings will be

saved. The display will return to the first screen of the Input Setup
Program.

If you do not want to save the settings, press the DOWN/NO button.
The display will return to the first screen of the Input Setup Program.

12. Press the ENTER button to perform additional setup or the ESCAPE

button to exit the Input Setup Program and begin the Beacon 200’s
warm-up sequence.

NOTE: Once the Beacon 200 channels are configured correctly, see the

detector head operator’s manual(s) for a complete description of
detector head installation procedures.