42 • Setting Up a New Channel or Changing an Existing Channel

If the detector head type is correct, press the UP/YES button to
proceed to the units and gas type screen and skip to step 7.

NOTE: See the detector head operator’s manual and the Beacon 200

Detector Head Specification sheet for the detector head to
determine the detector head type.

6. Use the UP/YES and DOWN/NO buttons to scroll through the list of

detector head types until the correct one is displayed. Press the
ENTER button to accept the type.The units and gas type screen
appears.

7. If the units and gas type are not correct, press the DOWN/NO button

and continue with step 8.

If the units and gas type are correct, press the UP/YES button to
proceed to the full scale screen and skip to step 9.

8. Use the UP/YES and DOWN/NO buttons to scroll through a list of units

and gas type choices.

NOTE: See the Beacon 200 Detector Head Specification Sheet for the

detector head to determine the correct units and gas type.

When the desired units and gas type appears, press the ENTER
button to accept the choice. The full scale screen appears.

One of the choices is User Will Specify. If the desired units and gas
type setting is not in the list, this setting will allow you to enter a 10
character units and gas type setting. With this choice displayed, press
the ENTER button and a screen will appear which allows you to input
the characters. Use the UP/YES and DOWN/NO buttons to select a
character and then press ENTER to accept it and continue with the
next character. When all characters have been entered, the full scale
screen will appear.

9. If the full scale setting is not correct, press the DOWN/NO button and

continue with step 10.

If the full scale setting is correct, press the UP/YES button to proceed
to the save screen and skip to step 11.

10. Use the UP/YES and DOWN/NO buttons to scroll through a list of full

scale settings.

NOTE: See the Beacon 200 Detector Head Specification Sheet for the

detector head to determine the correct full scale setting.

Advertising