Alarm indications – RKI Instruments Beacon 200 Rig Monitor, LEL/LEL User Manual

Page 27

Advertising
background image

Beacon 200 Rig Monitor Operator’s Manual • 22

Alarm Indications

This section describes the Rig Monitor in alarm 1, alarm 2, and fail
conditions and suggests response to these conditions. Table 5 below lists
the alarm indications for each condition.

NOTE: The Beacon 200 Rig Monitor includes alarm on and alarm off

delay settings for each channel and level of gas alarm. The alarm
indications described in this section operate according to the
factory set delay settings. See “Configure Channel Settings Menu”
on page 29 for all the factory settings.

Table 5: Visual and Audible Alarm Indications

Condition

Cause

Visual Indication(s)

Audible

Indication

Alarm 1

1

Increasing gas reading at or above the
alarm 1 setpoint

ALARM 1 light is on

Gas reading flashes and
alternates with ALARM-1
message

Strobe/horn turns on

Pulsing tone

Alarm 2

1

Increasing gas reading at or above the
alarm 2 setpoint

ALARM 2 light is on

Gas reading flashes and
alternates with ALARM-2
message

Pulsing tone

Fail

Disconnected or misconnected
detector wiring

Display reading below -10% of full
scale or lower

Defective components

FAIL light is on

FAIL message flashes in
place of gas reading

Steady tone

Low Battery

Dead 12 VDC battery.

FAIL light is on

SUPPLY VOLTAGE IS TOO
LOW
LOW POWER
STANDBY
message and
actual voltage of incoming
DC power

None

NOTE: Under typical operating conditions, there will be no low power alarm. The 12 VDC battery that is recommended for
operation will die before the power converter recognizes a low power condition.

*

1

If the Beacon 200 is in both an alarm 1 and an alarm 2 condition, both alarm lights are on and the display

alternates between the gas reading and the ALARM-1 ALARM-2 message.

Advertising