33 • Beacon 200 Rig Monitor Operator’s Manual

NOTE: The following procedure assumes that the target gas is present in

a high enough concentration to affect the fresh air (zero) reading.
If a fresh air environment can be verified, applying zero air to the
detector is not necessary when setting the zero reading.

This section describes how to calibrate the LEL detectors. It includes
procedures to prepare for calibration, set the fresh air reading, set the span
reading, and return to normal operation. It describes calibration using a
calibration kit that includes calibration gas, sample tubing, and a fixed flow
regulator with an on/off knob. RKI Instruments, Inc. recommends using a
0.5 LPM (liters per minute) fixed flow regulator. The calibration cup is part
of the detector assembly in the Rig Monitor.

Preparing for Calibration

1. Simultaneously press and hold the ENTER and UP/YES buttons for

approximately 5 seconds to enter the Calibration Program. Release the
buttons when the CALIBRATION PROGRAM... message displays and
asks if you want to continue or return to normal operation.

Calibrate

Channel 1 Y/N?

Calibration

Timeout Selection

ENTER

Normal Operation

Apply Gas to

Ch. 1 & Ch. 2

Detectors

Select Cal. Gas

Concentration for

Channel 2

Span

Channel 1 Y/N?

Span

Channel 2 Y/N?

Press Enter to

Adjust Span

Select Cal. Gas

Concentration for

Channel 1

Air Adjust Channel

1

Calibration

Program Enter

ENTER/ESCAPE

ESCAPE

Calibrate

Channel 2 Y/N?

Air Adjust Channel

2

Figure 10. Calibration Flow Chart