Beacon 200 Rig Monitor Operator’s Manual • 36

If you press the DOWN/NO button, the span of channel 1 will continue if
you selected it or if you did not select channel 1, the unit will return to
the first calibration program screen which asks if you want to continue
or escape from the Calibration Program.

3. If you pressed the UP/YES button for either channel, the display will

prompt you for the span gas concentration that will be used for the first
selected channel. The following instructions assume you pressed UP/
YES button for each channel.

4. Adjust the displayed concentration for channel 1 up or down as needed

using the UP/YES and DOWN/NO buttons so that it matches the
concentration in the calibration cylinder.

5. Press the ENTER button to accept the calibration gas concentration for

Channel 1 and continue.

6. The display will then prompt you for the span gas concentration that will

be used for channel 2. Adjust the displayed concentration up or down
as needed using the UP/YES and DOWN/NO buttons so that it matches
the concentration in the calibration cylinder.

7. Press the ENTER button to accept the calibration gas concentration for

Channel 2 and continue.

8. The unit will display the message Expose Dectector(s) To Gas. . .

When Done Press ENTER for a few seconds. It will then alternate
between the gas readings for both channels and the message
APPLYING GAS ENTER to ACCEPT ESCAPE to ABORT above
the time remaining in the calibration time-out.

9. Screw the calibration gas cylinder onto the fixed flow regulator.

10.Turn the regulator knob counterclockwise to open the regulator.

11.Allow calibration gas to flow for one minute.

12.Turn the regulator knob clockwise to close it. The Rig Monitor will

continue to display the maximum gas response on the display and
retain the response level in its memory.

13.Disconnect the sample tubing from the first detector and connect it to

the calibration cup hose barb of the second detector.

14.Turn the regulator knob counterclockwise to open the regulator.

15.Allow the calibration gas to flow for two minutes.

16.Turn the regulator knob clockwise to close the regulator. The Rig

Monitor will continue to display the maximum gas response on the
display and retain the response level in its memory.

17.Press the ENTER button at the Rig Monitor to proceed with the

calibration adjustment.