Specifications – RKI Instruments M2A User Manual

Page 7

Advertising
background image

M2A Transmitter Operator’s Manual

2

Specifications

Table 1 lists specifications for the M2A.

Table 1: M2A Specifications

Target Gas/Detection Range

Combustible Gas/Carbon Dioxide (CO

2

)

Catalytic LEL: 0 - 100% LEL, 1% LEL increments (CH

4

calibration unless otherwise

specified. H

2

specific detector available)

IR (infrared) CH

4

LEL: 0 - 100% LEL, 1% LEL increments

IR HC LEL: 0 - 100% LEL, 1% LEL increments (propane calibration)

IR CH

4

Volume: 0 - 100 % volume, 1% increments

IR CO

2

ppm: 0 - 5,000 ppm, 20 ppm increments

IR CO

2

Volume: 0 - 5.00 % volume, 0.01% increments

IR CO

2

Volume: 0 - 50.0 % volume, 0.1% increments

IR CO

2

Volume: 0 - 100 % volume, 1% increments

Oxygen

Oxygen (O

2

): 0 - 25.0 % volume, 0.1% increments

CO/H

2

S

Carbon Monoxide (CO): 0 - 300 ppm, 1 ppm increments

Hydrogen Sulfide (H

2

S): 0 - 100 ppm, 1 ppm increments

ESM-01 Toxic

Ammonia (NH

3

): 0 - 75.0 ppm, 0.1 ppm increments

Arsine (AsH

3

): 0 - 1.50 ppm, 0.01 ppm increments

Chlorine (Cl

2

): 0 - 3.00 ppm, 0.01 ppm increments

Chlorine (Cl

2

): 0 - 10.0 ppm, 0.1 ppm increments

Hydrogen Cyanide (HCN): 0 - 15.0 ppm, 0.1 ppm increments

Phosphine (PH

3

): 0 - 1.00 ppm, 0.01 ppm increments

Sulphur Dioxide (SO

2

): 0 - 6.00 ppm, 0.01 ppm increments

Alarm Settings
(Alarm 1/Alarm 2)

Combustible Gas (%LEL): 10 %LEL/50 %LEL
Combustible Gas (%Volume): 100 %/100 %
CO2 0 - 5,000 ppm: 2,500 ppm/5,000 ppm
CO2 0 - 5.00%: 0.50 %/3.00 %
CO2 0 - 50.0%: 50.0%/50.0%
CO2 0 - 100%: 100%/100%
Oxygen: 19.5%(decreasing)/23.5%(increasing)
Carbon Monoxide: 25 ppm/50 ppm
Hydrogen Sulfide: 10 ppm/50 ppm
Ammonia: 12 ppm/25 ppm
Arsine: 0.20 ppm/0.50ppm
Chlorine (0 - 3.00 ppm): 0.50 ppm/1.50 ppm
Chlorine (0 - 10.0 ppm): 1.0 ppm/3.0 ppm
Hydrogen Cyanide: 5.0 ppm/10.0 ppm
Phosphine: 0.10 ppm/0.30 ppm
Sulphur Dioxide: 1.00 ppm/2.00 ppm

Construction (housing)

Explosion-proof Junction Box, NEMA 4X

Area Classification

Explosion-proof for Class I, Groups B, C, and D (Combustible, CO

2

, CSA Type CO and

H

2

S, and CSA Type oxygen)

Sampling Method

Diffusion

Advertising