M2A Transmitter Operator’s Manual

WARNING:

When using the M2A, you must follow the instructions and warnings in this
manual to assure proper and safe operation of the M2A and to minimize the risk
of personal injury. Be sure to maintain and periodically calibrate the M2A as
described in this manual.

Input Power

10 - 30 VDC

Controls

Three push button switches

Three magnetic switches for non-intrusive calibration

Weight

4.5 lbs.

Signal Output

4 to 20 mA with impedance maximums listed below

-12 VDC input: 500 ohms impedance max
-24 VDC input: 1000 ohms impedance max

RS-485 Modbus

Explosion Proof

Non-Explosion Proof

Operating Temperature

LEL

-40°C to 75°C

N/A

Oxygen

-20°C to 45°C

-20°C to 50°C

H

2

S

-40°C to 40°C

-40°C to 50°C

CO

-5°C to 40°C

-20°C to 50°C

Toxic (ESM-01)

N/A

-10°C to 40°C

Infrared LEL

-20°C to 50°C

-40°C to 50°C

Infrared CO

2

-20°C to 50°C

-40°C to 50°C

Accuracy

Catalytic Combustible Gas:
± 5% of reading or ± 2% LEL (whichever is greater)
IR Combustible Gas/IR CO

2

:

± 5% of reading or ± 2% of full scale (whichever is greater)
Oxygen:
± 0.5% O

2

Hydrogen Sulfide:
± 5% of reading or ± 2 ppm H

2

S (whichever is greater)

Carbon Monoxide:
± 5% of reading or ± 5 ppm CO (whichever is greater)
ESM-01 Toxic Sensors:
± 10% of reading or ± 5% of full scale (whichever is greater)

*

These are the RKI factory settings. You can change the alarm settings in Configuration Mode. See

“Viewing & Changing M2A Parameters” on page 31.

Table 1: M2A Specifications