M2A Transmitter Operator’s Manual

Appendix A: Control Button Quick Reference Guide

The M2A’s control buttons allow access to operational modes, resetting of alarms, and display of
the Information Screen. Table 16 shows which button combinations perform these functions and
which parameters are available for adjustment while in the operational modes. While in these
modes, display prompts showing a “?” require you to respond by pressing either the UP/YES (for
yes) or DOWN/NO (for no) button. To change parameter settings, use the UP/YES and DOWN/NO
buttons to get the desired setting, then press and release the ENTER button to accept the setting.

Table 16: Control Button Quick Reference Chart

Mode Entered or

Function Performed

Button

Combination

Adjustable or Viewable Parameters

Alarm Reset

Press and Release
ENTER

n/a

Show Information Screen

Press and Hold
DOWN/NO

•

Input Voltage

•

M2A Version/Revision Information

Calibration Mode

Press and Hold UP/
YES for 5 seconds

•

Fresh Air Adjust

•

Span (or Zero for Oxygen) Adjust with Calibration
Gas

Configuration Mode

Press and Hold UP/
YES & ENTER for 5
seconds

•

Alarm Levels

•

Increasing or Decreasing Alarms

•

Normally Energized or Normally De-energized
Relays

•

Latching or Self-resetting Alarms

•

Alarm On Delay

•

Alarm Off Delay (if self-resetting)

•

Zero Suppression

•

Noise Filter

•

Calibration Time-out Period

Gas Type Mode

Press and Hold UP/
YES and DOWN/
NO for 5 seconds

Gas Type

Modbus Mode

Press and Hold
DOWN/NO and
ENTER for 5
seconds

•

Enable or Disable Modbus

•

Slave ID

•

Baud Rate

•

Parity

•

Response Delay