RKI Instruments GX-2009 User Manual
Page 29
Write Commands • 29
Follow these steps to calibrate an instrument:
1. Confirm that the area where you will perform the calibration
is a fresh air environment.
2. Launch the GX-2009 User Setup Program as described in
“Launching the Program” on page 14.
3. Connect the GX-2009 to the User Setup Program as
described in “Connecting a GX-2009 to the User Setup
Program” on page 15.
NOTE: It is not necessary to use Instrument Information or
Instrument Get & Save in order to perform a
calibration. Although it is not necessary to use
Instrument Information or Instrument Get & Save
before performing a calibration, you can perform a
calibration after using these either of these commands if
desired.
4. Click Calibration. The GX-2009 Calibration Window will
appear.
Figure 23: GX-2009 Calibration Window