Write Commands • 33

Figure 27: Installing the Adapter Plate, Auto Cal

10.Turn the regulator on/off knob counterclockwise to open it.

Calibration gas will begin to flow.

11.Allow the gas to flow for two minutes.

12.Click Span. The program will take a few seconds to adjust

the instrument’s span settings. The Adjusting Span Window
will appear during this period.