4.5 Linearity calibration
1.

Carry out a basic calibration, see above.

2.

Set the selected " minimum" temperatur e on a PT 100 simula-
tor (when using a pr ecision r esistance decade, wher e neces-
sary observe DIN IEC 751 conversion table from °C to

!)

and calibrate the signal conditioner output signal to 4.000 mA
using the null potentiometer .

3.

Set the mean temperature "(

ϑ

max

+

ϑ

min

)/2" on a PT 100 simu-

lator.
Determine the difference to the desired value:
"

I = shown value on measuring instrument - 12.000 mA"

and calibrate the output signal to (12.000 mA -

I) using a

span potentiometer .

4.

Set the " maximum" selected temperatur e

ϑ

min

on the PT 100

simulator and calibrate the signal output to 20.000 mA using
the lin potentiometer .

5.

Repeat steps 2 to 4 until the required accuracy is achieved.