Rockwell Automation 1440-VSE02-01RA XM-122 gSE Vibration Module User Manual
Page 196

Publication GMSI10-UM013D-EN-P - May 2010
186 Wiring Connections for Previous Module Revisions
Figure D.7 Powered Sensor to Channel 2 Wiring
Connecting Two Accelerometers and a Non-Contact Sensor
Figure D.8 shows the wiring of two IEPE accelerometers and a non-contact
sensor to an earlier revision of the XM-122 module (before revision D01). The
IEPE accelerometers are wired to channel 1 and channel 2. The non-contact
sensor is wired to the tachometer input signal.
1
17
38
+24V DC
Common
Signal
5
6
21
22
Channel 2 Input Signal
Signal Common
Shield
+24V DC
TYPICAL WIRING FOR MODEL 580 VIBRATION PICKUP
TO XM-122 VIBRATION MODULE CHANNEL 2
Cable shield not
connected at this end
ATTENTION
You may ground the cable shield at either end of the cable.
Do not ground the shield at both ends. Recommended
practice is to ground the cable shield at the terminal base
and not at the transducer. Any convenient Chassis terminal
may be used (see Terminal Block Assignments on page
179).
IMPORTANT
The buffered outputs are valid for all signals in this wiring
scheme.