4.13 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

102

■■■■■■■■

850 Professional IC –Anion – MCS – Prep 3

4.13.3.2

Cleaning the MSM

It may be necessary to clean the MSM in the following cases:

■

Increased backpressure onto the output capillaries of the MSM.

■

Blockage of the MSM which cannot be eliminated (solutions can no
longer be pumped through the MSM).

■

Jamming of the MSM which cannot be eliminated (MSM can no longer
be switched over).

1

2

3

4

5

Figure 54

MSM – Components

1

Union nut

2

MSM connecting piece 6.2832.010

3

MSM rotor A 6.2832.000

4

MSM housing

5

Slot in the MSM housing

Cleaning the MSM

Clean the MSM as follows:

1 Disconnecting the MSM from the IC system

■

Switch off the instrument.

■

Disconnect the MSM from the separation column, peristaltic
pump and detector.

2 Dismantling the MSM

■

Unscrew union nut (54-1) from the MSM housing (54-4).

■

Pull MSM connecting piece (54-2) and MSM rotor A (54-3) out of
the MSM housing (54-4). The MSM connecting piece and MSM
rotor A normally stick to one another - if this is not the case: Take
a sharp object, insert into the slot (54-5) in the MSM housing,
and pull out MSM rotor A (54-3) in this way.