Page 53: Injection and sequential suppression
Advertising

■■■■■■■■■■■■■■■■■■■■■■
2 Installation
850 Professional IC –Anion – MCS – Prep 3
■■■■■■■■
43
1
A
B
D
C
Figure 24
Cation exchange with the SPM
1
Sample
A
SPM
B
Injection valve
C
Separation column
D
MSM
2 Injection and sequential suppression
■
The injection valve is switched to the position Inject.
■
With the eluent flow (25-2) the sample prepared is rinsed onto
the separation column (25-C).
■
The sequential suppression with the MSM(25-D) and the MCS fol-
lows after the separation.
A
B
D
C
2
Figure 25
Injection and sequential suppression
2
Sample in the eluent flow
A
SPM
B
Injection valve
C
Separation column
D
MSM
Neutralization
A typical use of the SPM is the neutralization of an alkaline sample for
anion determination with chemical suppression.