The analyte ions are retained on the preconcentration column
(27-D) while the transfer solution flows to the waste container.
A
B
C
D
1
2
Figure 27
Neutralization and preconcentration
1
Sample
2
Transfer solution
A
Injection valve 1
B
SPM
C
Injection valve 2
D
Preconcentration column
3 Injection and sequential suppression
■
Injection valve 2 is switched:
– Injection valve 1 still is in position Inject.
– Injection valve 2 (28-C) in position Inject.
■
With the eluent flow (28-3) the analyte ions retained on the pre-
concentration column (28-D) are rinsed onto the separation col-
umn (28-E).
■
The sequential suppression with the MSM (28-F) and the MCS fol-
lows after the separation.
C
F
D
E
3
Figure 28
Injection and sequential suppression
3
Eluent
C
Injection valve 2