4 control parameters ep1, Control parameters ep1 – Metrohm 848 Titrino plus User Manual

Page 112

Advertising
background image

7.3 Endpoint titrations (SET)

■■■■■■■■■■■■■■■■■■■■■■

102

■■■■■■■■

848 Titrino plus

+
Positive measured value alteration, i.e. in the direction of a higher pH
value, greater voltage or greater current.

Negative measured value alteration, i.e. in the direction of a lower pH
value, lesser voltage or lesser current.
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.

Extraction time

Minimum duration of the titration. Titration will not be stopped during the
extraction time, even if the endpoint has already been reached. Titration is
however stopped if a stop condition is fulfilled during this time (see Chap-
ter 7.3.6, page 105)
. The entry of an extraction time may be advisable,
e.g. for the titration of sparingly soluble samples.

Input range

0 - 999999 s

Default value

0 s

Time interval MP

Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.

Input range

0.1 - 999999.0 s

Default value

2.0 s

7.3.4

Control parameters EP1

Menu

Parameters Control parameters EP1

The control parameters for the first endpoint are defined under Control
parameters EP1
.

Endpoint 1 at

Measured value for the first endpoint.

Measuring mode pH:
Input range

–20.000 - 20.000

Selection

off

Default value

off

Measuring mode U and Ipol:
Input range

–1250.0 - 1250.0 mV

Selection

off

Default value

off