9.2 Miscellaneous

■■■■■■■■■■■■■■■■■■■■■■

116

■■■■■■■■

848 Titrino plus

Problem

Cause

Remedy

The electrode responds too
slowly.

Replace the electrode.

9.2

Miscellaneous

Problem

Cause

Remedy

No report is printed.

The printer is not recog-
nized by the instrument.

■

Switch the 848 Titrino plus off and on
again.

■

Use a hub on the USB connector and con-
nect the printer to the hub.

The printer model is not
compatible.

Use a printer which fulfills the required specifi-
cations (see Chapter 10.4.4, page 121).

The USB keyboard
or the PC mouse
does not function.

The keyboard or mouse is
not recognized by the
instrument.

■

Switch the 848 Titrino plus off and on
again.

■

Use a hub on the USB connector and con-
nect the keyboard or mouse to the hub.

The keyboard or the mouse
is not compatible.

Use a model which fulfills the required specifi-
cations (see Chapter 10.4, page 119).

The display is no
longer readable.

The contrast is set incor-
rectly.

Adjust the contrast correctly (see Chapter 6.1,
page 52)
.

Mettler XP balances
send "R" or "O ----"
as ID1.

The automatic calibration
of the balance is switched
on.

Switch off the automatic calibration.

Message 020-511
"Action not possi-
ble" is displayed.

The USB stick is no longer
connected.

1. Connect the USB stick.
2. Switch the instrument off and on again.

The USB stick is full.

■

Use a different USB stick.

■

Delete files with the aid of a PC.