Metrohm 848 Titrino plus User Manual

Page 75

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 System settings

848 Titrino plus

■■■■■■■■

65

The interface parameters for the connected balance are set under COM1
settings
.

Baud rate

Transfer rate in characters per second.

Selection

1200 | 2400 | 4800 | 9600 | 19200 | 38400 |
57600 | 115200

Default value

9600

Data bits

Number of data bits.

Selection

7 | 8

Default value

8

Stop bits

Number of stop bits.

Selection

1 | 2

Default value

1

Parity

Type of parity testing.

Selection

even | none | odd

Default value

none

Handshake

Type of the data transfer protocol.

Selection

hardware | software | none

Default value

hardware

Note

If communications problems occur, set the parameter Handshake to
software, and make another attempt.

Editing the COM2 settings
Menu

▶ System ▶ External devices ▶ COM2 settings

The interface parameters for instruments connected to the RS-232/2 con-
nector of the RS-232/USB Box (e.g. PC) are set under COM2 settings.
The parameters and input ranges are identical to those for the COM1
interface.

Advertising