5 control, Control – Metrohm 848 Titrino plus User Manual

Page 39

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation

848 Titrino plus

■■■■■■■■

29

3 Exporting the method

■

In the function bar, select Export and press [OK].

The method is being exported. The directory structure on the USB
flash drive is listed in chapter 6.5, page 62.

5.5

Control

Menu

▶ Control

In the dialog Control, the settings for the execution of a single determi-
nation or of one sample series are defined.

Sample table

If this parameter is activated, the sample data for a sample series can be
entered in a table (see Chapter 5.7, page 32).

Selection

on | off

Default value

off

Autostart

If this parameter is activated, a new determination is started automatically
at the end of a determination. This continues until the number specified
has been reached (see Number of autostarts).

Selection

on | off

Default value

off

Number of autostarts

This parameter is visible only when Autostart = on.

Number of automatic starts.

Input range

1 - 50

Selection

table

Default value

table

table
The number of automatic starts corresponds to the number of samples
in the sample table.

Advertising