4.7 Results
■■■■■■■■■■■■■■■■■■■■■■
136
■■■■■■■■
873 Biodiesel Rancimat
Induction time
The induction time is used for extrapolation. If the induction time has been
determined manually, this value will be used and not the automatically
determined induction time.
Stability time
The stability time is used for extrapolation.
Formula 1 for induction time
The result calculated with formula 1 is used for extrapolation.
Formula 2 for stability time
The result calculated with formula 2 is used for extrapolation.
Formula 3 for stability time
The result calculated with formula 3 is used for extrapolation.
Target temperature
Temperature, for which the selected result is to be converted. The conversion
is realized using the regression straight line (from the extrapolation).
Range
freely selectable ... °C
"Live graph" tab
On the Live graph tab, you can carry out the settings for displaying the live
graph.