5 glp test set, 6 rs-232 interface, 7 mains connection – Metrohm 873 Biodiesel Rancimat Manual User Manual
Page 175: Glp test set, Rs-232 interface, Mains connection
Advertising

■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications
873 Biodiesel Rancimat
■■■■■■■■
167
7.5
GLP test set
External tempera-
ture sensor
Type
Pt100 casing element
⌀1.9 mm, in 4-conductor technology; DIN IEC 751
Class B
Calibration
accuracy
0…90 °C: ±0.02 °C
90…220 °C: ±0.03 °C
Max. deviation
of the measured
temperature
from the true
value (50…220
°C)
±0.3 °C
Test resistor
10 k
Ω
7.6
RS-232 interface
Plug
Sub-D plug, 9-pin (male)
Basic settings
9600 Baud, 8 Bit, 1 stop bit, no parity, XON/XOFF
7.7
Mains connection
Voltage
2.873.0014: 220…240 V ± 10 %
2.873.0015: 100…120 V ± 10 %
Frequency
50…60 Hz
Power consump-
tion
450 W
Fuse
Diameter 5 mm, length 20 mm
100…120 V: 2.0 ATH (slow-acting)
220…240 V: 4.0 ATH (slow-acting)
Advertising