Metrohm 873 Biodiesel Rancimat Manual User Manual
Page 70

4.5 Methods
■■■■■■■■■■■■■■■■■■■■■■
62
■■■■■■■■
873 Biodiesel Rancimat
Table 1
Approximate values for determining Delta T
Setpoint temperature
Delta T
80 °C
90 °C
100 °C
110 °C
120 °C
130 °C
140 °C
150 °C
160 °C
+0.7 °C
+0.8 °C
+0.8 °C
+0.9 °C
+1.0 °C
+1.1 °C
+1.1 °C
+1.2 °C
+1.2 °C
Gas flow
Gas flow through the sample. The gas flow can only be set for both blocks
jointly. The value of the method loaded for block A is always decisive.
Range
7 ... 25 L/h
Use internal pump
on | off (Default value: on)
Switch on/off gas supply with internal pump. If the internal pump is switched
off, an external gas supply must be connected to the Air/N
2
in connection
Start options
Start delay
Waiting time until starting the measurement.
Range
0 ... 1000 min
Start mode
Mode for starting and stopping the channels.
Selection
per channel | per 2 channels | per block
per channel
All channels can be started and stopped individually.
per 2 channels
2 channels each can be started and stopped jointly. The [Start] button
in the control window changes automatically for this.