3 glp status, Glp status – Metrohm 873 Biodiesel Rancimat Manual User Manual

Page 148

Advertising
background image

4.8 GLP functions

■■■■■■■■■■■■■■■■■■■■■■

140

■■■■■■■■

873 Biodiesel Rancimat

Tolerance

Indication of the tolerance value for the relevant GLP test. If the result is
within the range Setpoint ± Tolerance, the test is regarded as passed. If
you click on [Set default], the specified standard value is entered in this
field.

Additional alert message text

Additional message text which is to be added during the prompt for carrying
out the next GLP test.

4.8.3

GLP status

Displaying the GLP status

Proceed as follows to display the status of the GLP monitoring:

1 Open dialog window

If several instruments are connected: In the Control window, use
the menu item View

Unit x to select the required instrument.

Click on the menu item Tools

GLP test Status....

The dialog window GLP status for unit with serial no. x appears.

Parameter description

Test type

Type of GLP test (Temperature, Conductivity or Gas flow).

Next test due date

Date on which the next GLP test is to be carried out. If the GLP monitoring
is not switched on, Disabled appears in this column.

Due in

Number of days until the next GLP test.

Last test result

Indication whether the test has been passed (Ok) or not (Not Ok!).

Advertising