4 Operation
873 Biodiesel Rancimat
■■■■■■■■
53
■
Click on the symbol .
The dialog window with the method parameters is opened.
■
Under Temperature enter the required setpoint temperature for
the sample and set Delta T to 0 °C.
■
Set the gas flow to the required value and finalize with [Save].
5 Start gas flow and heating
■
Click on the menu item Tools
▶ Gas flow control ▶ Gas flow
on.
■
Click on the button [Heat to xx °C].
Executing a Delta T determination
After you have prepared everything for determining Delta T (see "Preparing
Delta T determination", page 52), you can now execute the determination
of Delta T as follows:
1 Enter sensor calibration data
■
Click on the menu item Tools
▶ Determine Delta T ▶ Block A
or Block B.
The following dialog window appears:
■
Enter a Sensor name.
■
Enter the calibration data (temperature and resistance) for Points
1 to 3 and confirm with [OK].