Calibration data, Solutions – Metrohm 757 VA Computrace User Manual

Page 116

Advertising
background image

5 Determination mode

757 VA Computrace – Software

106

Comments

Display of comments if any type of error appeared in the
sweep (e.g.

Ovl. in scan

= Overload during sweep;

Ovl. in

CDE

= Overload during cleaning or deposition;

Ovl. in

cond. cycles

= Overload during conditioning cycles;

No

peak found

= no peak found for defined substance;

Not

used

= peak is not used for calculation).

Calibration data

Substance

Substance name (see

Substance

on the

Substances

tab,

section 5.2).

Calibr.

Calibration technique

std.add.

,

rec.cc.

, or

smp.cc.

(see

Calibration

on the

Specifications

tab, section 5.4).

Y.reg/offset

Evaluation quantity for the sample calculated from the
standard addition curve (for standard addition) or inter-
cept of the calibration curve (for calibration curves).

Slope

Slope of the standard addition or calibration curve.

Mean deviat.

Calculated mean deviation of the measured values about
the standard addition or calibration curve.

Solutions

No.

Number of Dosimat used for addition or predose.

Content

Remarks regarding the solution (see

Content

on the

DO-

SIMATS

window, section 5.2).

Vol. (mL)

Volume of the exchange unit installed on the Dosimat
(see

Volume

on the

Hardware

tab, section 2.7).

Predose (mL)

If the Dosimat is used for solution predose, the message

Predose X.X mL

is displayed in this column.